Verify Class II Biosafety Cabinet laminar inflow/downflow pressure, sash alarm, UV-off safety, and 70% ethanol surface decontamination before work.
Turn off the UV germicidal lamp before opening the sash, switch on the blower and fluorescent light, and let the cabinet purge for at least 5–10 minutes.
Set the sash to the certified 8- or 10-inch working mark, confirm zero airflow alarms, and verify the HEPA pressure gauge sits within its green range.
Wipe the stainless steel work deck, side walls, and every item entering the BSC (pipettes, tip boxes, media bottles) with 70% ethanol or IPA.
Arrange supplies from clean-to-dirty zones at least 10 cm behind the front grille; never block front or rear intake grilles with wipes or arms.
Open flames disrupt the laminar airflow curtain, can damage the HEPA filter adhesive, and pose an explosion risk when combined with 70% ethanol vapors.
Purging allows the HEPA-filtered downflow air to sweep away settled particulates and establish a stable air barrier at the front sash opening.
Monitor ISO-classified cleanroom positive/negative differential pressures (Magnehelic gauges), temperature, and relative humidity per shift to prevent cross-contamination.
Verify chemical fume hood face velocity (80–120 fpm), sash working height stop, and baffle airflow before handling volatile or toxic reagents.
Perform daily level-bubble check, internal/external weight verification with Class E2/F1 NIST weights, and sensitivity check on analytical balances.