Audit laboratory and warehouse chemical storage for incompatible class segregation, secondary containment trays, flammable cabinet grounding, and GHS labels.
Confirm acids are separated from bases, oxidizers are isolated from flammables/organics, and water-reactive chemicals are stored in dry, sealed areas.
Verify all liquid corrosive and solvent bottles sit inside chemical-resistant secondary containment trays capable of holding 110% of the largest container.
Confirm yellow flammable cabinets are self-closing, grounded if dispensing, vented/capped per code, and not exceeding maximum allowable volume.
Ensure every bottle (including secondary aliquots) has a legible GHS hazard label, tight cap, and date-opened marking on peroxide-forming ethers.
Although both are acids, nitric acid is a strong oxidizer while acetic acid is an organic acid; mixing them can cause violent exothermic reactions or fire.
Opened peroxide formers like diethyl ether or THF should be dated upon receipt and opening, and tested or disposed of every 3 to 6 months per institutional EHS policy.
Monitor ISO-classified cleanroom positive/negative differential pressures (Magnehelic gauges), temperature, and relative humidity per shift to prevent cross-contamination.
Verify chemical fume hood face velocity (80–120 fpm), sash working height stop, and baffle airflow before handling volatile or toxic reagents.
Perform daily level-bubble check, internal/external weight verification with Class E2/F1 NIST weights, and sensitivity check on analytical balances.