Flush plumbed emergency eyewash stations and drench showers weekly to clear stagnant sediment, verify tepid water flow, and keep access paths clear.
Confirm the eyewash and shower are within 10 seconds (approx. 55 feet) of chemical hazards with zero carts, drums, or closed doors blocking the path.
Push the paddle/valve to verify hands-free operation, confirm dust caps pop off cleanly, and flush for 1–3 minutes until water is clear and tepid (16–38°C).
Place a test sock/bucket under the shower head, pull the triangular rod to confirm full-flow drench pattern, and verify the valve stays open hands-free.
Replace eyewash dust caps after flushing, check for leaks or corrosion, and sign the inspection log or scan the QR tag.
Weekly flushing clears stagnant water, rust scale, and opportunistic microorganisms like Acanthamoeba or Legionella from the supply pipes.
ANSI/ISEA Z358.1 defines tepid flushing fluid as 16°C to 38°C (60°F to 100°F) to prevent thermal shock or scalding during a full 15-minute flush.
Monitor ISO-classified cleanroom positive/negative differential pressures (Magnehelic gauges), temperature, and relative humidity per shift to prevent cross-contamination.
Verify chemical fume hood face velocity (80–120 fpm), sash working height stop, and baffle airflow before handling volatile or toxic reagents.
Perform daily level-bubble check, internal/external weight verification with Class E2/F1 NIST weights, and sensitivity check on analytical balances.