Verify ICH stability chamber temperature and humidity setpoints (25°C/60% RH, 40°C/75% RH), water reservoir levels, and alarm status daily.
Read both the chamber PLC controller and the independent calibrated data-logger display; verify temperature (±2°C) and humidity (±5% RH) are within ICH Q1A limits.
Inspect the RO/DI water supply tank for the humidity generator to ensure adequate level and zero algal or biofilm growth.
Confirm the chamber door seals tightly, audit-trail/recorder logging is active, and UPS/DG backup status shows normal.
If any alarm triggered over the past 24 hours, log the exact duration and magnitude and initiate a QA Deviation for ICH stability impact assessment.
Long-term stability for Zone II is typically 25°C ± 2°C / 60% RH ± 5% RH (or 30°C / 65% RH for Zone IVa), while accelerated testing is 40°C ± 2°C / 75% RH ± 5% RH.
Tap water contains dissolved minerals that cause scale buildup on steam heaters and ultrasonic transducers, leading to humidity control failures.
Monitor ISO-classified cleanroom positive/negative differential pressures (Magnehelic gauges), temperature, and relative humidity per shift to prevent cross-contamination.
Verify chemical fume hood face velocity (80–120 fpm), sash working height stop, and baffle airflow before handling volatile or toxic reagents.
Perform daily level-bubble check, internal/external weight verification with Class E2/F1 NIST weights, and sensitivity check on analytical balances.