Confirm hospital crash-cart defibrillators and public-access AEDs pass self-tests, hold full battery charge, and have unexpired electrode pads.
Inspect the AED or crash-cart monitor status window for a flashing green 'OK' or checkmark indicator and ensure no low-battery chirp is sounding.
For manual hospital defibrillators, select 30 Joules, charge with paddles seated in the test well, discharge, and verify the 'Test OK' printout.
Verify two sets of adult multifunction pads and one pediatric set/key are sealed, undamaged, and within their printed expiration date.
Ensure the razor, trauma shears, nitrile gloves, pocket mask, and conductive gel are in the pouch, and AC power is plugged into a red emergency outlet.
Hospital crash carts are high-acuity life-support devices used in acute cardiac arrests; NABH and Joint Commission standards mandate daily verification of charge and pad readiness.
Expired pads may have dried adhesive gel that causes skin burns or poor impedance reading. Always replace pads before their expiration date.
Standardize daily steam sterilizer Bowie-Dick air-removal tests, biological indicator spore monitoring, and cycle parameter verification in CSSD and clinics.
Monitor twice-daily minimum/maximum vaccine refrigerator temperatures between 2°C and 8°C and document immediate cold-chain excursion actions.
Verify ALS/BLS ambulance medical equipment, emergency drugs, oxygen pressure, and vehicle safety systems at every shift handover.