Standardize daily steam sterilizer Bowie-Dick air-removal tests, biological indicator spore monitoring, and cycle parameter verification in CSSD and clinics.
Before the first cycle, clean the chamber drain strainer of lint or debris, wipe the silicone door gasket, and verify the steam generator water level.
Place a Bowie-Dick test pack horizontally on the bottom rack over the drain in an empty pre-vacuum sterilizer and run the 134°C / 3.5-minute test cycle.
Place wrapped surgical trays on edge without touching chamber walls or stacking flat, ensuring chemical indicator strips are visible inside and outside each pack.
Review the physical printout for minimum 134°C (or 121°C gravity) hold time and pressure, then incubate the Geobacillus stearothermophilus biological indicator.
Allow packs to cool to room temperature on wire racks; reject any wet or torn packs for immediate re-sterilization before signing the load release log.
Per AAMI ST79 and NABH guidelines, a Bowie-Dick test must be run every day the pre-vacuum steam sterilizer is used, prior to processing the first patient load.
A wet pack compromises the sterile barrier through wicking and must be treated as contaminated. Do not release it; re-wrap and re-sterilize the entire pack.
Geobacillus stearothermophilus spores are the standard biological indicator for steam autoclaves because of their high resistance to moist heat.