Verify ALS/BLS ambulance medical equipment, emergency drugs, oxygen pressure, and vehicle safety systems at every shift handover.
Start the ambulance engine, test emergency beacons, siren, and cabin inverter power, and verify the main wheeled stretcher locks securely into the floor antler.
Open main D/M-type oxygen valves briefly to confirm pressure is ≥1000 psi (or >50% capacity), check regulators for leaks, and verify adult/pediatric masks.
Power on the portable suction unit to confirm ≥300 mmHg vacuum, verify defibrillator battery charge and pad expiry, and run a glucometer strip check.
Confirm tamper-evident seals on the ALS drug box and airway kit; if broken, restock against the par list and check all medications for 30-day expiry.
Ambulances should undergo a full readiness check at the start of every shift (daily or every 12 hours) and immediately after every patient transport.
Both the onboard main cylinder and portable cylinder should be at least 50% full (typically ≥1000 psi on a 2000 psi tank) to ensure uninterrupted supply during long transfers.
Standardize daily steam sterilizer Bowie-Dick air-removal tests, biological indicator spore monitoring, and cycle parameter verification in CSSD and clinics.
Monitor twice-daily minimum/maximum vaccine refrigerator temperatures between 2°C and 8°C and document immediate cold-chain excursion actions.
Confirm hospital crash-cart defibrillators and public-access AEDs pass self-tests, hold full battery charge, and have unexpired electrode pads.