Inspect clinical laboratory centrifuges for rotor balance, lid-interlock integrity, aerosol-containment seals, and RPM tachometer verification.
Remove the rotor and buckets; check for hairline cracks, corrosion, or dried specimen spills, and verify O-rings on aerosol-tight safety cups.
Confirm the centrifuge cannot start with the lid unlatched and remains locked until the rotor comes to a complete standstill.
Check RPM using an optical tachometer (or annual calibration sticker) against set speed (tolerance ±5%) and verify timer accuracy with a stopwatch.
Wipe the bowl with 70% isopropyl alcohol or non-corrosive disinfectant, dry thoroughly, and verify balance tube sets are available beside the unit.
Keep the lid closed for at least 30 minutes to allow infectious aerosols to settle before wearing PPE and using forceps to remove broken glass.
Relative Centrifugal Force (RCF or ×g) accounts for rotor radius, ensuring consistent sample separation across different centrifuge models.
Standardize daily steam sterilizer Bowie-Dick air-removal tests, biological indicator spore monitoring, and cycle parameter verification in CSSD and clinics.
Monitor twice-daily minimum/maximum vaccine refrigerator temperatures between 2°C and 8°C and document immediate cold-chain excursion actions.
Verify ALS/BLS ambulance medical equipment, emergency drugs, oxygen pressure, and vehicle safety systems at every shift handover.